884X-RTD產(chǎn)品概述
884X-RTD 100 Ohm RTD Temperature Probe
Features:
Manufactured using a coil suspension element design for increased shock and vibration resistance
Mineral-insulated sheath with a minimum bend radius of 19 mm (3/4 in) for flexibility and durability
Additional Specs:
Resistance: Nominal 100 Ω
Temperature coefficient: 0.00385 Ω/Ω …C nominal
Temperature range: -200 C to 300 …C (transition and cable temperature: 0 …C to 25% …C)
Drift rate: + 0.13 …C at 0 …C after 1000 hours at 300 …C
Sheath Material: 316 stainless steel
Leads: PTFE-insulated, nickel-plated stranded copper, 22 AWG
Termination: 4-wire banana
Time constant: Four seconds maximum for 63.2% response to step change in water moving at 3 fps
Bending radius: 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip
Calibration: Includes manufacturer’s NIST-traceable calibration and table with R vs. T values in 1 …C increments from -196 …C to 300 …C. Callendar-van Dusen coefficients included
Immersion: At least 100 mm (4 in) recommended
Accuracy (includes calibration uncertainty and short-term stability):
±0.50 …C at -196 …C
±0.050 …C at 0 …C
±0.052 …C at 200 …C
±0.055 …C at 300 …C
Size: 9 in L x 3/16 in diameter
Marking: Each RTD is marked with the Fluke name and part number